Author: Kroeze H. Van Vulpen M. De Leeuw A. A. C. Van De Kamer J. B. Lagendijk J. J. W.
Publisher: Informa Healthcare
ISSN: 1464-5157
Source: International Journal of Hyperthermia, Vol.17, Iss.3, 2001-05, pp. : 240-257
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
Local pain is the main factor that limits regional hyperthermia treatment. Using the SAR model of the regional hyperthermia treatment planning system, the capability of absorbing blocks to reduce peripheral hot spots was investigated. The effect of rectangular absorbers of various size and salinity on an elliptical phantom in the Coaxial TEM was evaluated. The computed results were compared with SAR values measured in the phantom. Absorbers of 9×9×4cm3 and a salinity of 18graml-1 provide a SAR reduction in the muscle equivalent material, centrally under the absorber of at least 50% at a depth of up to 3cm. The effect on the central (i.e. tumour) region is less than 20%. Larger absorbers have a more global effect and cause more attenuation in the central region. The attenuating effect depends strongly on the thickness of the fat layer between muscle and absorber. More than 2cm fat limits the effective use of absorbers. Absorbers can induce a significant increase of SAR in muscle and fat near their edges. This effect also depends on absorber size and salinity and the thickness of the fat layer. The effect of an absorber was also evaluated with a patient anatomy, yielding results in agreement with the phantom experiments.
Related content






By Cho Chi-Hee Sreenivasa Geetha Plotkin Michail Pietsch Hubertus Wust Peter Lüdemann Lutz
International Journal of Hyperthermia, Vol. 26, Iss. 4, 2010-09 ,pp. :


By Wielheesen D. H. M. de Bruijne M. Graveland W. J. van Rhoon G. C. van der Zee J.
International Journal of Hyperthermia, Vol. 21, Iss. 1, 2005-02 ,pp. :