Author: McDonald John T. Williams J. C. Williams James C.
Publisher: Society for Applied Spectroscopy
ISSN: 0003-7028
Source: Applied Spectroscopy, Vol.43, Iss.4, 1989-05, pp. : 697-702
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
This paper describes the evaluation of a time-resolved graphite spark source emission spectrometer built for the purpose of determining Ca, Na, and K in nanoliter volume samples of renal fluid. Subnanogram detection limits calculated from the working curve slope and the uncertainty of the sample signal are reported for Ca, Na and K; these detection limits are some three orders of magnitude higher than the shot noise limit. With the use of Li as an internal standard, working curves from 0 to 1 ng of Na, K, and Ca are shown. System performance data are given, and the utility of the analysis is discussed.