Author: Jun J. W. Lee H. S. Kwon T. Y. Minogin V. G.
Publisher: IOP Publishing
ISSN: 0026-1394
Source: Metrologia, Vol.38, Iss.3, 2001-06, pp. : 221-227
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
An evaluation of the light shift in an optically pumped caesium-beam frequency standard is presented. We show that the light shift can be accurately estimated by taking into account the angular anisotropy of the spontaneous photon emission, the neglect of which is shown to give rise to an error comparable to the value of the shift. Our data refer to a complete set of optical transitions for the caesium D2 line relevant to the KRISS-1 caesium-beam frequency standard.
Related content






Accuracy evaluation of the primary frequency standard NIST-7
By Shirley J. H. Lee W. D. Drullinger R. E.
Metrologia, Vol. 38, Iss. 5, 2001-10 ,pp. :


Accuracy evaluation of the primary frequency standard NIST-7
By Shirley J. H. Lee W. D. Drullinger R. E.
Metrologia, Vol. 39, Iss. 1, 2002-02 ,pp. :


A 633 nm iodine-stabilized diode-laser frequency standard
By Edwards C. S. Barwood G. P. Gill P. Rowley W. R. C.
Metrologia, Vol. 36, Iss. 1, 1999-01 ,pp. :