3D conformal MRI-controlled transurethral ultrasound prostate therapy: validation of numerical simulations and demonstration in tissue-mimicking gel phantoms

Author: Burtnyk Mathieu   N'Djin William Apoutou   Kobelevskiy Ilya   Bronskill Michael   Chopra Rajiv  

Publisher: IOP Publishing

ISSN: 0031-9155

Source: Physics in Medicine and Biology, Vol.55, Iss.22, 2010-11, pp. : 6817-6839

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

MRI-controlled transurethral ultrasound therapy uses a linear array of transducer elements and active temperature feedback to create volumes of thermal coagulation shaped to predefined prostate geometries in 3D. The specific aims of this work were to demonstrate the accuracy and repeatability of producing large volumes of thermal coagulation (>10 cc) that conform to 3D human prostate shapes in a tissue-mimicking gel phantom, and to evaluate quantitatively the accuracy with which numerical simulations predict these 3D heating volumes under carefully controlled conditions. Eleven conformal 3D experiments were performed in a tissue-mimicking phantom within a 1.5T MR imager to obtain non-invasive temperature measurements during heating. Temperature feedback was used to control the rotation rate and ultrasound power of transurethral devices with up to five 3.5 × 5 mm active transducer elements. Heating patterns shaped to human prostate geometries were generated using devices operating at 4.7 or 8.0 MHz with surface acoustic intensities of up to 10 W cm−2. Simulations were informed by transducer surface velocity measurements acquired with a scanning laser vibrometer enabling improved calculations of the acoustic pressure distribution in a gel phantom. Temperature dynamics were determined according to a FDTD solution to Pennes' BHTE. The 3D heating patterns produced in vitro were shaped very accurately to the prostate target volumes, within the spatial resolution of the MRI thermometry images. The volume of the treatment difference falling outside ±1 mm of the target boundary was, on average, 0.21 cc or 1.5% of the prostate volume. The numerical simulations predicted the extent and shape of the coagulation boundary produced in gel to within (mean ± stdev [min, max]): 0.5 ± 0.4 [−1.0, 2.1] and −0.05 ± 0.4 [−1.2, 1.4] mm for the treatments at 4.7 and 8.0 MHz, respectively. The temperatures across all MRI thermometry images were predicted within −0.3 ± 1.6 °C and 0.1 ± 0.6 °C, inside and outside the prostate respectively, and the treatment time to within 6.8 min. The simulations also showed excellent agreement in regions of sharp temperature gradients near the transurethral and endorectal cooling devices. Conformal 3D volumes of thermal coagulation can be precisely matched to prostate shapes with transurethral ultrasound devices and active MRI temperature feedback. The accuracy of numerical simulations for MRI-controlled transurethral ultrasound prostate therapy was validated experimentally, reinforcing their utility as an effective treatment planning tool.

Related content