Enzymes Immobilized on a Tubular Reactor for Fast Amperometric Determination of Glucose in Brazilian Honey

Author: Silva Vanezia   Cerqueira Marcos Rodrigues Facchini   Matos Maria Auxiliadora Costa   Matos Renato Camargo  

Publisher: Taylor & Francis Ltd

ISSN: 0003-2719

Source: Analytical Letters, Vol.42, Iss.1, 2009-01, pp. : 154-165

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

Glucose present in honey was rapidly determined by the differential amperometric method using two tubular reactors containing glucose oxidase and peroxidase. The linear dynamic range extends from 5 × 10-5 to 2 × 10-4mol L-1, at pH 7.0. At flow rate of 1.5 mL min-1 and injecting 150-µL sample volumes, a sampling frequency of the 33 determinations per hour is afforded. The reproducibility of the methods showed a relative standard deviation (RSD) < 4%. The detection limit of this method is 1.7 × 10-5 mol L-1. The samples analyses were compared with the parallel spectrophotometric determination.