Implantation of Novel Small-Diameter Polyurethane Vascular Prostheses Interposed in Canine Femoral and Carotid Arteries

Publisher: Karger

E-ISSN: 1421-9921|36|4|241-248

ISSN: 0014-312x

Source: European Surgical Research, Vol.36, Iss.4, 2004-07, pp. : 241-248

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

Objectives: The performance of small-diameter (3–5-mm) vascular grafts still poses a challenge in the field of vascular surgery. We present here our preliminary experience with implanting unique small-sized polycarbonate urethane vascular grafts in 7 dogs. Material and Methods: Each animal was implanted with 4 interposition grafts, 2 femoral and 2 carotid. No anti-thrombotic medication was administered. Doppler sonography was performed at 3-month intervals to examine for patency and flow characteristics. Animals were sacrificed electively at 3, 6 and 12 months. Results: At 3 months, all grafts were patent. After 6 months, 3 grafts occluded and at 1 year a further 6 grafts occluded. Hence 9 of 28 grafts occluded (67.9% patency). During the study, no correlation could be established between flow velocity or resistance index and occlusion. Histopathology showed intimal hyperplasia to be the cause of occlusion. Conclusions: Compared to literature data on small-diameter grafts in the same position, ADIAM’s Biomechanical grafts performed clearly better. Compliance data suggest a correlation between elastic compliance and patency.