Author: MOORE C S BEAVIS A W SAUNDERSON J R
Publisher: British Institute of Radiology
ISSN: 0007-1285
Source: British Journal of Radiology, Vol.81, Iss.970, 2008-10, pp. : 771-777
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
The purpose of this study was to determine the optimum tube voltage and amount of added copper (Cu) filtration for processed chest radiographs obtained with an Agfa 75.0 Computed Radiography (CR) system. The contrast-to-noise ratio (CNR) was measured in the lung, heart/spine and diaphragm compartments of a validated chest phantom using various tube voltages and amounts of Cu filtration. The CNR was derived as a function of air kerma at the CR plate and with the effective dose. As rib contrast can interfere with detection of nodules in chest radiography, a tissue-to-rib ratio (TRR) was derived to investigate which tube voltages suppress the contrast of rib. Although processing algorithms affect the signal and noise in a way that is hard to predict, we found that, for a given set of processing parameters, the CNR was related to the plate air kerma and effective dose in a logarithmic manner (all