Performance of KCl:Eu 2+ storage phosphor dosimeters for low-dose measurements

Author: Li H Harold   Xiao Zhiyan   Hansel Rachael   Knutson Nels   Yang Deshan  

Publisher: IOP Publishing

ISSN: 0031-9155

Source: Physics in Medicine and Biology, Vol.58, Iss.12, 2013-06, pp. : 4357-4366

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

Recent research has demonstrated that europium doped potassium chloride (KCl:Eu2+) storage phosphor material has the potential to become the physical foundation of a novel and reusable dosimetry system using either film-like devices or devices similar to thermoluminescent dosimeter chips. The purposes of this work are to quantify the performance of KCl:Eu2+ prototype dosimeters for low-dose measurements and to demonstrate how it can be incorporated into clinical application for in vivo peripheral dose measurements. Pellet-style KCl:Eu2+ dosimeters, 6 mm in diameter, and 1 mm thick, were fabricated in-house for this study. The dosimeters were read using a laboratory photostimulated luminescence detection system. KCl:Eu2+ prototype storage phosphor dosimeter was capable of measuring a dose-to-water as low as 0.01 cGy from a 6 MV photon beam with a signal-to-noise ratio greater than 6. A pre-readout thermal annealing procedure enabled the dosimeter to be read within an hour post-irradiation. After receiving large accumulated doses (∼10 kGy), the dosimeters retained linear response in the low-dose region with only a 20% loss of sensitivity comparing to a fresh sample (zero Gy history). The energy dependence encountered during low-dose peripheral measurements could be accounted for via a single point outside-field calibration per each beam quality. With further development the KCl:Eu2+−-based dosimeter could become a versatile and durable dosimetry tool with large dynamic range (sub-cGy to 100 Gy).