Detection of high incidence of Riata lead breaches by systematic postero-anterior and lateral chest X-ray in a large cohort

Author: Steinberg Christian   Sarrazin Jean-François   Philippon François   Bouchard Marc-André   O'Hara Gilles   Molin Franck   Nault Isabelle   Blier Louis   Champagne Jean  

Publisher: Oxford University Press

ISSN: 1099-5129

Source: EP-Europace, Vol.15, Iss.3, 2013-03, pp. : 402-408

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

AimsInsulation breaches with externalization of conductor cables have been described for St-Jude Medical Riata™ defibrillation leads. Published data on the incidence of Riata lead abnormalities are quite heterogeneous. The objective of this study was to estimate systematically the prevalence of lead abnormalities using a postero-anterior (PA) and lateral chest X-ray (CXR).Methods and resultsFrom 2002 to 2008, 552 Riata defibrillation leads were implanted at our centre. We evaluated patients for potential insulation breaches. A PA and lateral CXR was obtained. Chest X-rays were reviewed by two electrophysiologists using a zooming function with magnification up to factor 7.5 and were classified as normal or abnormal for the presence of conductor externalization. A total of 284 patients were included. Riata lead models were 1570, 1580, 1582, 1590, 1592, 7000, 7002, and 7022. The total frequency of radiological lead defects was 24.3%. Insulation breaches occurred at zones of major lead curvature. Mean maximal spacing between extruding lead components was 3.6 ± 1.9 mm (range 2.0-12.4). Abnormal CXRs were more frequent in 8F leads (31.4% vs. 6.3%; P < 0.001). Most defects occurred with lead models 1582 (41.2%) and 1580 (31.4%). Mean time since implantation was longer in abnormal leads (6.7 vs. 5.9 years; P < 0.001). Abnormal leads had higher pacing thresholds (1.1 ± 0.8 V vs. 0.9 ± 0.4 V; P = 0.02).ConclusionThe incidence of insulation breach in Riata leads is much higher than quoted by the manufacturer or reported by most of the literature. A PA and lateral CXR with zooming appears adequate to identify lead breaches when reviewed by an electrophysiologist. Riata lead breaches without electrical abnormalities present a management dilemma and will require further studies.