QRS Detection Based on Morphological Filter and Energy Envelope for Applications in Body Sensor Networks

Author: Zhang Fei  

Publisher: Springer Publishing Company

ISSN: 1939-8018

Source: Journal of Signal Processing Systems, Vol.64, Iss.2, 2011-08, pp. : 187-194

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

Emerging body sensor networks (BSN) provide solutions for continuous health monitoring at anytime and from anywhere. The implementation of these monitoring solutions requires wearable sensor devices and thus creates new technology challenges in both software and hardware. This paper presents a QRS detection method for wearable Electrocardiogram (ECG) sensor in body sensor networks. The success of proposed method is based on the combination of two computationally efficient procedures, i.e., single-scale mathematical morphological (MM) filter and approximated envelope. The MM filter removes baseline wandering, impulsive noise and the offset of DC component while the approximated envelope enhances the QRS complexes. The performance of the algorithm is verified with standard MIT-BIH arrhythmia database as well as exercise ECG data. It achieves a low detection error rate of 0.42% based on the MIT-BIH database.