Assessment of Endothelial Function Using Arterial Pressure Signals

Author: Wu Hsien-Tsai  

Publisher: Springer Publishing Company

ISSN: 1939-8018

Source: Journal of Signal Processing Systems, Vol.64, Iss.2, 2011-08, pp. : 223-232

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

Peripheral arterial tonometry (PAT) is a useful method to assess endothelial function. Analysis of pulse wave amplitude with PAT during reactive hyperemia can be used to study peripheral vascular endothelial function very well. We had injected a specific nitric oxide (NO) synthesize inhibitor intra-arterially into five adult male rats to verify the validation of our proposed photoplethysmograpy (PPG) system for endothelial function assessment. However, the reproducibility was not good for the system applied for measuring fingertip peripheral arteries. Therefore, an air-pressure sensing system with the high reproducibility was developed for peripheral vascular endothelial function assessment. This study showed that the vasodilatation index measured with air-pressure sensing system correlated very well with that measured using peripheral arterial tonometry. In addition, only few minutes are needed for conducting a self endothelial function assessment at home. And, early self-monitoring of cardiovascular dysfunction and arterial stiffness can be easily and effectively achieved.