Author: Rosser S. de la Mora J.
Publisher: Taylor & Francis Ltd
ISSN: 1521-7388
Source: Aerosol Science and Technology, Vol.39, Iss.12, 2005-12, pp. : 1191-1200
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
A variant of the Vienna DMA (Winklmayr et al. 1991; Reischl et al. 1997) with inner and outer electrode radii of 25 and 33 mm, and a conventional trumpet inlet diameter of 97 mm has been tested. It incorporates a reduced pressure drop sheath gas exhaust system that enables reaching flow rates approaching 4000 L/min. Several new additional flow features are included to delay the transition to turbulent conditions. A cylindrical geometry with a DMA length L of 97 mm (distance between the aerosol inlet and outlet slits) is seen to keep the flow laminar up to the highest Reynolds number achieved, though showing slight signs of flow quality deterioration at about Re = 20,000. An equally long DMA with an inner electrode shaped as a 5° cone caped by a spherical dome remains stable up to the highest Reynolds number achieved. It is expected to continue this trend to considerably higher flow rates. Both these long configurations exhibit line widths close to the ideal Brownian diffusion limit, reaching FWHH of 4% for a particle mass diameter of 1 nm. A short DMA with an axial length L of 18 mm and an inner electrode shaped as a 5° cone caped by an ellipse of 3/2 aspect ratio remains also laminar at the highest speeds attained, and exhibits FWHH as small as 2.4%. It departs moderately from ideal behaviour at small flow rates, presumably due to its non-cylindrical geometry. More serious departures observed at high Reynolds numbers may perhaps be due to flow unsteadiness radiated into the working section by sound waves from the turbulent exhaust region. This is the first report of a DMA capable of excellent resolution at 1 nm, yet with a sufficiently wide and long working section to enable (in principle) covering the size range up to 100 nm.
Related content


A Medium Flow, High-Resolution Vienna DMA Running in Recirculating Mode
By Steiner G.
Aerosol Science and Technology, Vol. 44, Iss. 4, 2010-04 ,pp. :


Effect of Acoustic Radiation on DMA Resolution
Aerosol Science and Technology, Vol. 39, Iss. 9, 2005-09 ,pp. :


High density plug-type conveying
Advanced Powder Technology, Vol. 2, Iss. 1, 1991-01 ,pp. :