Abbott Propoxyphene Assay: Evaluation and Comparison of TDx FPIA and GC/MS Methods

Author: Kintz Pascal   Mangin Patrice  

Publisher: Oxford University Press

ISSN: 0146-4760

Source: Journal of Analytical Toxicology, Vol.17, Iss.4, 1993-07, pp. : 222-224

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

This study evaluated the capability of the Abbott TDx assay to test for propoxyphene in urine and various biological samples, including tissues obtained from three fatal overdoses, by comparison to gas chromatography/mass spectrometry (GC/MS). First, within-run and between-run precision were determined using three control samples (200, 400, and 900 ng/mL) tested over a two-week period. Within-run coefficients of variation (CV) for the three controls were 1.4, 2.2, and 2.5%, respectively; the between-run CVs were 2.5, 3.1, and 4.0%, respectively. The cross-reactivity with norpropoxyphene, the major metabolite of propoxyphene, was concentration dependent and in the range of 29.3 to 92.6%. Propoxyphene and its metabolite were assayed in biological samples the same day using the Abbott TDx and GC/MS. Tissue preparations were analyzed by TDx without specimen pretreatment other than homogenization and dilution with saline. The TDx results were in accordance with the results obtained by GC/MS.