Author: Dumas Pierre Gingras Guy LeBlanc Alain
Publisher: Oxford University Press
ISSN: 0146-4760
Source: Journal of Analytical Toxicology, Vol.29, Iss.1, 2005-01, pp. : 71-75
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
A direct and sensitive method for the determination of blood cyanide by isotope dilution was developed. The blood is placed in a headspace vial, and K13C15N is added as internal standard. Addition of phosphoric acid liberates the cyanide as HCN. The detection is accomplished by mass spectrometry after a fine mass calibration tuning. The detection limit obtained is 0.3 µmol/L. The within- and inter-run coefficients of variation are 4.4% (for a concentration of 2.5 µmol/L) and 3.9% (for a concentration of 4.7 µmol/L), respectively. The observed recovery is 98%. A round-robin exercise was carried out to compare the performance of this method with others currently in use in other clinical laboratories.