Author: Cormier Maurice Thompson James L.
Publisher: Society for Applied Spectroscopy
ISSN: 0003-7028
Source: Applied Spectroscopy, Vol.24, Iss.4, 1970-07, pp. : 465-466
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
There appears to be no widely used method of measuring the temperatures of liquid samples in ir spectroscopy, although Potts has suggested the use of a resistance thermometer. Other authors have used thermocouples, but common types of thermocouples will not fit into the sample spaces of ir liquid cells so that a nagging doubt remains about the actual temperature of the sample in the beam. In this investigation, temperatures inside the cell were determined with an accuracy of ± 0.1°C by means of a thermocouple made from two 0.0025 cm diam Platinel wires spot welded together.