Author: Mathews Samuel E. Ekimoff David Walters John P.
Publisher: Society for Applied Spectroscopy
ISSN: 0003-7028
Source: Applied Spectroscopy, Vol.36, Iss.6, 1982-11, pp. : 617-626
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
A hardware system is described which provides for software-independent timing of events necessary for the smooth operation of an adjustable waveform spark source. Illustrative cases are given showing the extension of this system into a single-channel spectrometer and a complex schlieren photography experiment. Software outlines are also provided. The use of an optical coupler and a fiber optic are shown to improve system performance greatly in high radio frequency interference environments.
Related content


Historical Advances in Spark Emission Spectroscopy
Applied Spectroscopy, Vol. 23, Iss. 4, 1969-07 ,pp. :


Infrared Emission Spectroscopy. I. Basic Considerations
Applied Spectroscopy, Vol. 26, Iss. 1, 1972-01 ,pp. :