Evaluation of Recycling Cyclone Spray Chambers for ICP-AES

Author: Vieira Peter A.   Zhizhuang He   Chan Shi-Kit   Montaser Akbar  

Publisher: Society for Applied Spectroscopy

ISSN: 0003-7028

Source: Applied Spectroscopy, Vol.40, Iss.8, 1986-11, pp. : 1141-1146

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

Two cyclone spray chambers of different sizes are evaluated for inductively coupled plasma/atomic emission spectrometry. On the basis of a simplified mathematical model, the cut-off diameters of the aerosol droplets are estimated to be 2.2 to 4.3 and 1.8 to 3.5 μm for the large and the small cyclones, respectively. The sample consumption rate can be reduced to about 50 μL/min when the cyclone chamber is used in a recycling mode. In general, the signal-to-background ratios, detection limits, and precisions of the analyte signal intensities obtained with the small cyclone chamber are slightly superior to those achieved with a Scott spray chamber and a recycling gravitational sedimentation chamber. The performance of the recycling cyclone chamber is also evaluated for the determination of copper in the NBS freeze-dried urine.