Author: Niemczyk T. M. Gobeli G. W.
Publisher: Society for Applied Spectroscopy
ISSN: 0003-7028
Source: Applied Spectroscopy, Vol.42, Iss.5, 1988-07, pp. : 792-796
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
A direct-device, i.e., no sine bar, grating drive has been built and characterized. The use of a microprocessor controller allows the system to be calibrated easily. Once calibrated the system demonstrates precision and accuracy that are as good as theoretically possible given the gearing and stepper motor selected. The lack of a sine bar mechanism also allows the grating to be rotated to very high angles. Wavelength scans with the use of a single grating are demonstrated.
Related content






By Lee Ted Killeen Owen P. MacIntyre S. A.
Applied Spectroscopy, Vol. 15, Iss. 4, 1961-07 ,pp. :