Author: Victor Mark A. Crouch Stanley R.
Publisher: Society for Applied Spectroscopy
ISSN: 0003-7028
Source: Applied Spectroscopy, Vol.49, Iss.7, 1995-07, pp. : 1041-1047
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
Synchronous fluorescence spectroscopy is frequently used to resolve multiple components in mixtures without separation. However, uncompensated inner-filter effects can lead to unexpected and/or large quantitative errors. An instrument and a method that correct for primary and secondary inner-filter effects in synchronous fluorescence spectroscopy have been designed and characterized. A bifurcated fiber-optic-based diode array fluorometer that simultaneously measures front-surface fluorescence and absorbance was constructed. A transfer function that mathematically describes the coupling efficiency between the excitation and emission legs of the fiber optics is defined and characterized. Once characterized, such a fluorometer is capable of calculating what the fluorescence measurement would be in the absence of inner-filter effects. Several experimental procedures for determining the instrument's transfer function are compared and discussed. Applications of the various procedures to experimental fluorescence inner-filter corrections to problematic samples are demonstrated.