Author: Kovaleva S. Gladyshev V. Rubinskaya T.
Publisher: MAIK Nauka/Interperiodica
ISSN: 1061-9348
Source: Journal of Analytical Chemistry, Vol.61, Iss.1, 2006-01, pp. : 72-76
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
A procedure is proposed for the voltammetric determination of selenium as selenosulfate (SO3Se2−) ions at a mercury-film electrode (MFE). Selenosulfate ions are determined in the range from 2 × 10−4 to 1.0 × 10−3 M without analyte accumulation, using peak current at −0.92 ± 0.02 V and in the range from 1 × 10−7 to 2 × 10−4 M after analyte accumulation with the open circuit, using peak current at −1.18 ± 0.03 V as the analytical signal. The mechanisms of SO3Se2− reduction at an MFE under the conditions of direct voltammetry and stripping voltammetry with accumulation are proposed and discussed.