Characterization of a dynamic multi-leaf collimator for stereotactic radiotherapy applications

Author: Godwin G A   Simpson J B   Mugabe K V  

Publisher: IOP Publishing

ISSN: 0031-9155

Source: Physics in Medicine and Biology, Vol.57, Iss.14, 2012-07, pp. : 4643-4654

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

The Apex® dynamic mini-multileaf collimator has recently been released by Elekta and attaches directly to the linear accelerator head. This paper details the work and results obtained in characterizing this mini-MLC for stereotactic usage within our department. A range of mechanical and dosimetric characteristics were investigated which included inter and intra leaf leakage, light/radiation field congruence, leaf position reproducibility, radiation penumbra, total scatter factors and mechanical rotational stability with the additional mini-MLC weight.

Related content