A Deviation Function Used for the Secondary Realization of the ITS-90 in the Subrange from 83.8033 K to 273.16 K

Author: Kang Zhiru   Lan Jingbo   Zhang Jintao   Sun Jianping   Jin Chen  

Publisher: Springer Publishing Company

ISSN: 0195-928X

Source: International Journal of Thermophysics, Vol.31, Iss.8-9, 2010-09, pp. : 1484-1490

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

This article proposes a correlation relation between the resistance ratios of the triple points of argon and mercury. By this relation, the resistance ratio of the triple point of argon can be extrapolated from that of mercury, and a deviation function which is defined in the range from 83.8058 K to 273.16 K can be determined from only the calibration values at the triple points of water and mercury. It is a close approximation to the ITS-90 deviation function in the subrange. Using it, the calibration at the triple point of argon can be saved. Twenty-five standard platinum resistance thermometers are used to check the function. The errors are less than 5 mK. It is sufficient for secondary measurements.