Radiochemical sensor system for the analysis of 99Tc(VII) in groundwater

Author: Hartman J.  

Publisher: Springer Publishing Company

ISSN: 0236-5731

Source: Journal of Radioanalytical and Nuclear Chemistry, Vol.264, Iss.2, 2005-05, pp. : 495-500

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

Direct determination of 99Tc in groundwater using a sensor device represents a substantial challenge due to the short range of beta-particles, the presence of radiological interferences, and low-level detection requirements. This paper describes a radiochemical sensing method for selective detection of 99Tc(VII) in water. The sensing approach uses a dual-function composite sensor microcolumn that incorporates both selective sorption and scintillating properties. Analyte detection is carried out in a reagentless equilibration sensing regime using chemically unmodified groundwater. The sensor method was implemented in a prototypical probe device that integrates the sensor element, scintillation detection, data acquisition, and instrument control components in a single functional instrument compatible with a 3.5" well casing geometry. The feasibility of rapid 99Tc analysis in Hanford groundwater (Hanford Site, USA) below regulatory drinking water level of 33 Bq/l was demonstrated.