Author: Buehler Christof Kim Ki Greuter Urs Schlumpf Nick So Peter
Publisher: Springer Publishing Company
ISSN: 1053-0509
Source: Journal of Fluorescence, Vol.15, Iss.1, 2005-01, pp. : 41-51
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
We present a multicolor multiphoton fluorescence microscope with single-photon counting sensitivity. The system integrates a standard multiphoton fluorescence microscope, an optical grating spectrograph operating in the UV–Vis wavelength region, and a 16-anode photomultiplier tube (PMT). The major technical innovation is in the development of a multichannel photon counting card (mC-PhCC) for direct signal collection from multi-anode PMTs. The electronic design of the mC-PhCC employs a high-throughput, fully-parallel, single-photon counting scheme along with a high-speed electrical or fiber-optical link interface to the data acquisition computer. There is no electronic crosstalk among the detection channels of the mC-PhCC. The collected signal remains linear up to an incident photon rate of 108 counts per second. The high-speed data interface offers ample bandwidth for real-time readout: 2 MByte λ-stacks composed of 16 spectral channels, 256× 256 pixel image with 12-bit dynamic range can be transferred at 30 frames per second. The modular design of the mC-PhCC can be readily extended to accommodate PMTs of more anodes. Data acquisition from a 64-anode PMT has been verified. As a demonstration of system performance, spectrally resolved images of fluorescent latex spheres and
Related content


By Zhang Chunling Zhang Liying Yang Ru Liang Kun Han Dejun
Applied Spectroscopy, Vol. 67, Iss. 2, 2013-02 ,pp. :




By Kleeblatt Juliane Ehlert Sven Hölzer Jasper Sklorz Martin Rittgen Jan Baumgärtel Peter Schubert Jochen K. Zimmermann Ralf
Applied Spectroscopy, Vol. 67, Iss. 8, 2013-08 ,pp. :