Determination of Low-Level Residual Ethylene Oxide by Using Solid-Phase Microextraction and Gas Chromatography

Author: Ayoub Kamal   Harris Leonard   Thompson Bill  

Publisher: AOAC International

ISSN: 1944-7922

Source: Journal of AOAC International, Vol.85, Iss.6, 2002-11, pp. : 1205-1209

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

Current methods of analysis for ethylene oxide (EO) in medical devices include headspace and simulated-use extractions followed by gas chromatography with either a packed or a capillary column. The quantitation limits are about 0.5–1.0 μg/g for a packed column and about 0.1–0.2 μg/g for a capillary column. The current allowable levels of EO on medical devices sterilized with EO gas as outlined in International Organization for Standardization (ISO) 10993-7 may be significantly reduced from current levels by applying the ISO Draft International Standard 10993-17 method for establishing allowable limits. This may require EO test methods with detection and quantitation limits that are much lower than those of the currently available methods. This paper describes a new method that was developed for the determination of low-level EO by solid-phase microextraction using the direct-immersion method. Factors such as temperature and stirring were found to affect absorption efficiency and absorption time. A low extraction temperature (about 6°C) was found to be more efficient than room-temperature extraction. Stirring was found to reduce absorption time by about 50%. Under these conditions, detection and quantitation limits of 0.002 and 0.009 μg/g, respectively, were obtained by using a capillary column. As a result, this method makes compliance with lower EO limits feasible.

Related content