Abstract
Raman optical activity (ROA) has been exclusively observed in the visible (VIS) and near‐infrared (NIR) spectral regions to date. During the last few years, we have designed, constructed and tested the first ROA instrument, operating in the deep‐ultraviolet (DUV) spectral region employing 244‐nm excitation. This novel DUV ROA instrument is based on a backscattering geometry and incident circular polarization modulation (ICP); it makes use of a fast DUV imaging lens‐based spectrograph and specially designed DUV grade polarization optics. The performance of this instrument has been evaluated by analysing measured non‐resonant DUV ROA spectra of non‐absorbing enantiomeric liquid samples and by comparing these with corresponding ROA spectra recorded in the visible spectral region. Copyright © 2015 John Wiley & Sons, Ltd.