In non-polluted natural waters, trace metals have generally to be determined near the detection limit of electrothermal atomic absorption spectrometry (ETAAS) and therefore the experimental conditions must be optimized. The detection limit in ETAAS depends strongly on numerous experimental parameters and its optimization is time consuming because, according to IUPAC recommendations, 20 repeated blank measurements per experimental parameter must be performed. In this work, for lead the influence of different experimental parameters on the instrumental detection limit together with the relatively good agreement between experimental and