Author: Moreno E. Corns W. T. Bryce D. W. Stockwell P. B.
Publisher: Taylor & Francis Ltd
ISSN: 1464-5068
Source: Journal of Automated Methods & Management in Chemistry, Vol.22, Iss.2, 2000-03, pp. : 33-39
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
The procedure developed allows the direct speciation of arsenic in these samples with good sensitivity, selectivity, precision and accuracy. Detection limits determined using the optimized conditions were found to be between 0.16 and 2.9ng ml1 for arsenite, dimethylarsinic acid, monomethylarsonic acid and arsenate, while standard addition studies showed that the procedure is free from matrix interferences. As no certified reference materials are available for these analytes or matrices, validation was carried out by studying spike recoveries and by comparison of results with an alternative technique.
Related content




JOURNAL OF SEPARATION SCIENCE (ELECTRONIC), Vol. 38, Iss. 15, 2015-08 ,pp. :


JOURNAL OF SEPARATION SCIENCE (ELECTRONIC), Vol. 38, Iss. 6, 2015-03 ,pp. :


JOURNAL OF SEPARATION SCIENCE (ELECTRONIC), Vol. 37, Iss. 15, 2014-08 ,pp. :


CHINESE JOURNAL OF CHEMISTRY (ELECTRONIC), Vol. 33, Iss. 2, 2015-02 ,pp. :