Author: Turbino Míriam
Publisher: Springer Publishing Company
ISSN: 0268-8921
Source: Lasers in Medical Science, Vol.26, Iss.4, 2011-07, pp. : 421-425
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
The aim of this study was to define optimal power settings as well as curing time associated with evaluating the curing depth of a composite resin as a function of Vickers hardness. The tests were performed with a hybrid composite resin cured with a halogen lamp and argon ion laser, with different exposure times and power settings. The composite resin bulk technique was used using a black polypropylene matrix with thicknesses ranging from 1 to 4 mm and Vickers microhardness was measured on the opposite surface of the light activation. ANOVA and Tukey statistical tests were used. The results showed that the groups activated by the laser for 20 s, at 200 and 250 mW, did not present statistically significant differences regarding the halogen lamp with 1 mm thickness, but the halogen lamp showed better results with thickness values more than 2 mm (