Author: Nakayama Kenji
Publisher: Springer Publishing Company
ISSN: 1053-0509
Source: Journal of Fluorescence, Vol.21, Iss.2, 2011-03, pp. : 835-839
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
We propose a conclusive difference observed between the excitation conditions required to observe porphyrins and copper-metallothioneins in cells and/or tissues using an ordinary fluorescence microscope. We have emphasized the importance of examining the spectral properties of the emissions to avoid any serious mistakes such as confusing porphyrins with copper-metallothioneins in the liver and kidneys. However, microspectrophotometry is not a conventional method for either histochemical, cytochemical, or pathological studies because microspectrophotometers are both expensive and difficult to operate. Therefore, we demonstrate a simple comparative method using ordinary excitation filter arrangements. When using our technique, it becomes possible to optically discriminate more accurately between the autofluorescence properties arising from porphyrins and those arising from copper-metallothioneins. We would like to name our simple technique “Triple Observation Method (TOM).