Author: Palmisano Pietro Accogli Michele Zaccaria Maria Luzzi Giovanni Nacci Frida Anaclerio Matteo Favale Stefano
Publisher: Oxford University Press
ISSN: 1099-5129
Source: EP-Europace, Vol.15, Iss.4, 2013-04, pp. : 531-540
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
AimsThe long-term impact of implantable device-related complications on the patient outcome has not been thoroughly evaluated. The aims of this retrospective, bi-centre study were to analyse the rate and nature of device-related complications requiring surgical revision in a large series of patients undergoing device implantation, elective generator replacement and pacing system upgrade and to systematically assess the impact of such complications on patient outcome and healthcare utilization.Methods and resultsData from 2671 consecutive procedures (1511 device implantations, 1034 elective generator replacements, and 126 pacing system upgrades) performed between January 2006 and March 2011 were retrospectively analysed. The outcome measures recorded were complication-related mortality, number of re-operations, need for complex surgical procedures, number of re-hospitalizations, and additional hospital treatment days. Over a median follow-up of 27 months, the overall rate of complications was 2.8% per procedure-year [9.5% in cardiac resynchronisation therapy (CRT) device implantation, 6.1% in pacing system upgrade, 3.5% in implantable cardioverter defibrillator implantation, 1.7% in pacemaker implantation, and 1.7% in generator replacement). The procedure with the highest risk of complications was CRT device implantation (odds ratio: 6.6;