On the operational quantity H p (3) for eye lens dosimetry

Author: Behrens R.  

Publisher: IOP Publishing

ISSN: 0952-4746

Source: Journal of Radiological Protection, Vol.32, Iss.4, 2012-12, pp. : 455-464

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

In the past the operational quantity Hp(3) has been defined for calibration purposes in a slab phantom. Recently, an additional phantom in the form of a cylinder was suggested for eye lens dosimetry as a cylinder much better approximates the shape of a head than a slab. Therefore, this work investigates whether the quantity Hp(3), when defined in the respective calibration phantom, adequately estimates the eye lens dose (or is at least conservative) depending on the phantom: it turns out that in most cases both calibration phantoms are similarly well suited. Finally, the definition of the eye lens dose is discussed together with possible consequences on the definition of Hp(3): the consideration of only the radiation sensitive volume of the lens causes Hp(3) not to be conservative in beta radiation fields.