Flow Injection Chemiluminescence Determination of Phenol in Neutral Medium

Author: Hao Liang   Du Jianxiu   Lu Jiuru  

Publisher: Taylor & Francis Ltd

ISSN: 0003-2719

Source: Analytical Letters, Vol.44, Iss.1-3, 2011-01, pp. : 38-47

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

A new flow injection chemiluminescence method for the determination of phenol was proposed, based upon the chemiluminescence reaction of phenol, N-bromosuccinimide, and hydrogen peroxide in neutral aqueous medium in the presence of cetyltrimethylammonium bromide surfactant micelles. The chemiluminescence signal was proportional to the concentration of phenol in the range of 1.0 × 10-7-8.0 × 10-6 g/mL with a detection limit of 3 × 10-8 g/mL. The relative standard deviation for 1.0 × 10-6 g/mL phenol solution was 2.0% (n = 11). The proposed method was successfully applied to the determination of phenol in phenol ear drops. A possible CL reaction mechanism was also discussed briefly.