Development and Validation of an Analytical Method for Quantification of Arsenic and Antimony in Liposomes Using Inductively Coupled Plasma-Optical Emission Spectrometry

Author: Reis Priscila G.   Abreu Adriana T.   Guimarães Andrea G.   Teixeira Mônica C.   de Souza Jacqueline   Silva-Barcellos Neila M.  

Publisher: AOAC International

ISSN: 1944-7922

Source: Journal of AOAC International, Vol.96, Iss.4, 2013-07, pp. : 771-775

Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.

Previous Menu Next

Abstract

Arsenic and antimony compounds are used to treat endemic diseases, such as cancer, leishmaniasis, and schistosomiasis, in spite of their toxicity. Several studies seeking the development and characterization of nanocarrier systems such as liposomes are being carried out with the aim of developing new drug delivery systems and minimizing the toxicity of these drugs. However, the lack of reference methods to quantify these semimetals within a liposomal matrix hinders the QC of these formulations. Therefore, the validation of an analytical method for arsenic and antimony quantification in liposomal matrix by inductively coupled plasma-optical emission spectrometry is presented here. The linearity, specificity, LOD, LOQ, accuracy, and precision were determined according to the International Conference on Harmonization norms and the Brazilian Health Surveillance Agency (Resolution 899). The LOD values were 0.02 and 0.06 mg/L for antimony and arsenic, respectively. The LOQ for both was 3.0 mg/L, with an adequate accuracy within 98.26 and 101.32% for different levels of antimony and 99.98 and 100.36% for arsenic. Precision (CV) was lower than 5.0%. The developed and validated method was shown to be reproducible for quantification of arsenic and antimony in liposome pharmaceutical dosage forms.

Related content