Author: Delahaye F. Witt T. J. Pesel E. Schumacher B. Warnecke P.
Publisher: IOP Publishing
ISSN: 0026-1394
Source: Metrologia, Vol.34, Iss.3, 1997-09, pp. : 211-213
Disclaimer: Any content in publications that violate the sovereignty, the constitution or regulations of the PRC is not accepted or approved by CNPIEC.
Abstract
Abstract. An on-site comparison of the quantum Hall effect (QHE) resistance standards of the PhysikalischTechnische Bundesanstalt (PTB) and of the Bureau International des Poids et Mesures (BIPM) was made in October 1995. Measurements of a 100 Omega standard in terms of R(H)2, the quantized resistance corresponding to the quantum number 2, agreed to -5,5 parts in 1010 with a relative combined standard uncertainty uc=19x10-10. Measurements of 10 000 Omega/100 Omega and 100 Omega/1 Omega ratios agreed to-4 parts in 1010 with uc = 19x10-10 and 12 parts in 1010 with uc=26 10-10, respectively.
Related content






Direct comparison of cryogenic radiometers from the BIPM and the PTB
By Goebel R. Pello R. Stock K. D. Hofer H.
Metrologia, Vol. 34, Iss. 3, 1997-09 ,pp. :


On the synthesis of quantum Hall array resistance standards
Metrologia, Vol. 52, Iss. 1, 2015-02 ,pp. :


Comparison of the Josephson voltage standards of the IEN and the BIPM
By Reymann D. Witt T. J. Andreone D. Cerri R. Godone A.
Metrologia, Vol. 35, Iss. 1, 1998-02 ,pp. :